• Lotix/ LSS Boat

Lotix/ LSS Boat


TOC users from around the globe requested simplicity, performance, and durability. The Teledyne Tekmar LSS Boat Module is an ideal add-on to the Lotix Total Organic Carbon (TOC) analyzer. It uses high-temperature combustion to accurately measure carbon content in a variety of matrices: soils, sludges, sediments, particulate-laden liquids and hard-to-oxidize samples.

Compact and simple, the LSS Boat is designed to work with the Lotix making it the ultimate solution for all of your TOC needs both liquid and solid for years of operation with an economical price to fit any budget.

Key Features:
  • Automated sample delivery into furnace via software
  • Low cost of ownership
  • Up to 1 gram sample size giving you a wide analytical range
  • Ability to switch back and forth between liquids and solids within minutes
  • Analysis of 1 gram samples with up to 20mg carbon content can reduce errors associated with using small sample sizes​

Chemistry:

Oxidation by Combustion: From 680 °C - 1000 °C

Analytical Method:

TC, TOC, NPOC

Detector:

Non-Dispersive Infrared (NDIR) Detector - requires Lotix TOC Analyzer

Boat Material:

Quartz

Controller:

PC, Interface through Windows™ 7 (with SP1) - Professional, Enterprise, and Ultimate (32 or 64 bit), Windows™ 10 Professional Enterprise, or greater

TOC Analytical:

S.D. +/- 2% typical of a mid-range standard

Maximum Sample Amount:

Up to 1.5 grams depending on sample matrix

Measuring Range:

TC, TOC 0.1 mg to 20 mg carbon

TOC Analysis Time:

~5-7 minutes with a gas rate of 200 mL/min

Sample Introduction:

Automated delivery of sample boat into furnace via software

Principal Applications:

Soils, sludges, sediments, particulate-laden liquids, hard-to-oxidize samples

Certification:

CE (CSA site certified if required)

TOC Utility Requirements:

Voltage: 100/115/230 VAC (±10%) (Operation on nominal 100 VAC mains input requires a minimum of 108V achieved by the addition of a step-up transformer with a minimum 2000 Watts (VA) power rating.), Frequency: 50/60 Hz

TOC Dimensions:

10” (25.4 cm) W x 24” (60.9 cm) D x 15.8” (40.1 cm) H Weight 32 lbs (14.5 kg)

TOC Gas Supply:

Hydrocarbon and Carbon Dioxide (CO2) free air with TOC content <1 ppm or UHP O2. Gas can be supplied from a cylinder or TOC gas generator. If a TOC gas generator is used, resulting gas must be hydrocarbon and water free.

LSS Gas Inlet Flow:

200 mL/min from Lotix +/- 10%